Beschreibung
V8016 Weight Indicator
1.Introduction:
The instrument adopts 24-bit A/D converter, which is combined with various sensors and transmitter to measure, display, monitor, collect and record the physical parameters such as pressure, flow rate, material position, component analysis, force and mechanical quantity.
2. Technical parameters
Work environment | Temperature: 0-50℃ | Indicator: | Double 5-bit high brightness red LED |
Moisture:20-90%RH | Output voltage: | 5v±5%(supply for load sensor),current<50mA | |
Accuracy | ±0.2%F.S | Supply voltage: | 195~242 Vac, Power<5w;Or 24Vdc |
Measurement cycle time | 0.2s |
3. The key introduction
“Set”: set and confirm the parameter “>”: Shift key or peeling key (for peak meter)
“∧”:add the data “CLR”: Peak reset key or peeling key (for up and down limit instrument)
4.Dimensions
Panel size | |
L | H |
160mm | 80mm |
5.Parameters description
Symbol | Name | 范 围 |
SA1 | OUT1 relay output | 00000~full scale |
SA2 | OUT2 relay output | 00000~ full scale |
SA3 | OUT3 relay output | 00000~ full scale |
SA4 | OUT4 relay output | 00000~ full scale |
SL | Calibration“0” | 00000 |
SH | Load calibration | Above half full scale |
Sd | Display decimal places selected | 0~3 |
Id | Division value | 0.1,0.2,0.5,1,2 |
CLTF | Tare key selection | 0.1 |
FIL | Weight filtering | 1 to 9 |
OUTF | Alarm output selection | 0 or 1, 0 is alarm when less setting; 1 is alarm when above setting. |
6. Operating instructions
6.1 Press “SET” and hold 3 seconds, display “SA1”, press “>”to set out1 relay output value; press “SET”, display “SA2” to set out2 relay output value; press “SET”, display “SA3” to set out3 relay output value; press “SET”, display “SA4” to set out4 relay output value;
6.2 Press “SET”, display “CIN”, Password input, setting “00000” is valid. If you want to keep the password, press “SET”, to set next; if you want to change the password, press “CLR”, display “CON”, to input new password, press “SET”, to set next;
6.3 Press “SET” display “SL”, to set the unload Zero Calibration.
6.4 Press “SET” display “SH”, to set the load calibration (must above half full scale), input the load value.
6.5 Press “SET” display “Sd”, to set Display decimal places, press “∧” to change the decimal places.
6.6 Press “SET” display “Id”, to set Division value, you can press “∧” to choose 0.1,0.2,0.5,1,2.
6.7 Press “SET” display “CLTF”, to set Tare key selection, when “CLTF” is “0”, press “CLR” can tare to the indicator; when “CLTF” is “1”, press “CLR” can’t tare to the indicator.
6.8 Press “SET” display “FIL”, to set Weight filtering, Choose 1-9, the larger the value, the slower the display. This value affects stability. Press “SET” to work station.
6.9 Press “SET” display “OUTF”, to set Alarm output selection, when the 5th digit tube displays “1”, OUT1 relay output signal when the value is above “OUT1” setting value; when the 5th digit tube displays “0”, OUT1 relay output signal when the value is less than “OUT1” setting value; set the 4th,3rd,2nd digit displays to set OUT2, OUT3, OUT4 by the same way.
6.10 Indicator connecting instruction:
Connecting load cell or junction box IN+/IN-:11-green,12-white
Connecting load cell or junction box E+/E-:9-Red,10-Black
Connecting LED data cable.23-blue,24-brown
GND
Supply power
7. Notes
When the indicator display “E1”, it means no connecting the load sensor or the signal of load sensor is too big.
8. Emergency
8.1 When forgot the password, press “CLR” and power on, the password will be recover “00000”.
8.2 When mis-set or the button doesn’t work, press “∧”and power on, the indicator restore factory defaults.
8.3 Press “>” and power on, the digital tube changes between 0 and 1, when display “0”, loose the key, the indicator will be Upper and lower limit instrument; when display “1”, loose the key, the indicator will be peak value instrument;